Trust is good. Measuring is better!
Designation: wall mountable device for trace humidity / moisture traces

NEW!!! Lower detection limit 0.01 ppmV | 10 ppbV

The wall mounted device AQUATRACE® - IV  Typ W belongs to the group of AQUATRACE® - IV devices that serve as instruments for determination of traces of moisture / humidity traces.

They operate on the basis of P2O5 sensors and can be used for determining the moisture in corrosive and all other gases (except NH3) in the ppm range. The use in chlorine gas has been proven worldwide best.   

The wall mounted units with IP 65 degree of protection (standalone) can be integrated directly in industrial plants. It is as a 1 or 2-channel device available.

Special features of the moisture meters of type AQUATRACE® - IV

  • The operation of the Moisture meters ensues via a touchscreen display
  • in 7 languages to be select
  • Menu-driven operation and for various operating levels protected by password
  • Measured values and operator actions are logged internally and can be read out via the USB interface
  • The event logs are in CSV format as a daily file available
  • The devices for measurement of traces of moisture / humidity traces are executed according with the latest NAMUR recommendations
  • Production is carried out in close exclusive cooperation with the MSC Ltd electronic components, Grossroehrsdorf according to the latest industry standards

Additionally there are the following versions for your needs

  • The AQUATRACE 19" – version is available as a 1-, 2-, 3-, or 4-channel device
  • Laboratory device (portable version) is also available as a two-channel device.

 

 

Technical specifications

Product:

AQUATRACE® IV - Typ W, wall mount device, 1-2 channels

 

Performance specifications

Measuring range:

0-2000 Vppm, dew point –100 ° C - 20 ° C, ranges deviating from the standard on request

Accuracy:

+/- 5% of the displayed value, at least 0.4 ppmV | optional 0.04 ppmV

Response time:

dry to wet: T90 <5 sec; Wet to dry: T90 <15 min

Repeatability

+/- 0.1 ppmV | optional +/- 0.02 ppmV

Calibration

traceable calibration certificate, n-point calibration/adjustment (optional)

Display

Touchscreen 800x480 pixels, color

 

Every measuring channel

Lower detection limit

0.1 ppmV | optional: 0.01 ppmV

Measuring range

0-2000 Vppm

Displayed value(s)

In ppmV or Tp ° C (adjustable)

Transmission rate

32 bit

Data log rate

min. 2 s max 10 min

Sensor detection

automatically

Automatic sensor test

Automatically (fuction test)

Analog output

4-20 mA (freely scalable)

Alarm output

limit value - programmable as floating changeover contact

Channel relay

4 channel relays according to NAMUR (F + S; C + M), per channel (max. 2)

 

Measuring cell

Material measuring cell

PVDF or stainless steel

sensor material

Platinum on glass / ceramic or interdigital structure on ceramic substrate

Leakage rate

< 5x10 -7 mbar x l / s

Sample Gas flow

20 Nl/h or 100 Nl/h cell

Sample gas temperature

+5 … +90 °C

Sample gas pressure

approx. 200 mbar overpressure at the measuring cell inlet, pressureless flow out

Gas connections

¼“ NPT inside thread

Cell connection (el.)

5-pin Binder plug, angled

Measuring cable

Customized 2 m; Max. Cable length 100 m

« Back to overview
This email address is being protected from spambots. You need JavaScript enabled to view it.