The AQUATRACE® - IV Type - A can be used either as a battery-operated measuring device for mobile trace moisture measurement in gases or with power supply. AQUATRACE® - IV-family is supplemented by the AQUATRACE® -IV Type - A, a portable cordless tool with integrated power supply. After separation from the power, the device can work at least 2 hours without external power source. Assuming that the sensor coating is carried out in network operation

The AQUATRACE® - IV Type T location-independent and flexible to use. Measuring device for mobile trace moisture measurement in gases. AQUATRACE® - IV-family is supplemented by the AQUATRACE® -IV Type T, a portable Analyzer but without battery.

All the known options of AQUATRACE® - IV are at the same levels of service.

 

 

 

 

 

Technical specifications

Product:

AQUATRACE® IV - Typ A/T, mobile device with optional battery, 1-2 channels

 

Performance specifications

Measuring range:

0-2000 Vppm, dew point –100 ° C - 20 ° C, ranges deviating from the standard on request

Accuracy:

+/- 5% of the displayed value, at least 0.4 ppmV | optional 0.04 ppmV

Response time:

dry to wet: T90 <5 sec; Wet to dry: T90 <15 min

Repeatability

+/- 0.1 ppmV | optional +/- 0.02 ppmV

Calibration

traceable calibration certificate, n-point calibration/adjustment (optional)

Display

Touchscreen 800x480 pixels, color

 

Every measuring channel

Lower detection limit

0.1 ppmV | optional: 0.01 ppmV

Measuring range

0-2000 Vppm

Displayed value(s)

In ppmV or Tp ° C (adjustable)

Transmission rate

32 bit

Data log rate

min. 2 s max 10 min

Sensor detection

automatically

Automatic sensor test

Automatically (fuction test)

Analog output

4-20 mA (freely scalable)

Alarm output

limit value - programmable as floating changeover contact

Channel relay

4 channel relays according to NAMUR (F + S; C + M), per channel (max. 2)

 

Measuring cell

Material measuring cell

PVDF or stainless steel

sensor material

Platinum on glass / ceramic or interdigital structure on ceramic substrate

Leakage rate

< 5x10 -7 mbar x l / s

Sample Gas flow

20 Nl/h or 100 Nl/h cell

Sample gas temperature

+5 … +90 °C

Sample gas pressure

approx. 200 mbar overpressure at the measuring cell inlet, pressureless flow out

Gas connections

¼“ NPT inside thread

Cell connection (el.)

5-pin mini Binder plug, angled

Measuring cable

Customized 2 m; Max. Cable length 100 m