Trust is good, but measuring is even better!

The AQUATRACE® easyLINE is our new, particularly inexpensive, P2O5 trace moisture analyzer for ppmV | ppbV for inert and combustible gases with integrated sampling system. The system measures the volume-related moisture content in the gas and outputs it as ppmV and also approximately converts this value into the corresponding dew point temperature. Suitable for process and industrial moisture measurement, the measuring system is also equipped with limit alarm relays and status relays. The 1.8" TFT display visualizes the measured value. If the range is exceeded or not reached, the display color changes to flashing red. In addition to an analogue output 4-20mA, the analyzer also has an RS232 interface, which can be used to read out the measured value, but also for adjustment to local conditions.

Particularly suitable for the following sectors:

  • Glove boxes | insulators
  • Vacuum technology
  • Plasma, sintering and soldering furnaces
  • Technologies with protective gases (welding; soldering; thermal treatments, etc.)
  • Coating plants

Further advantages of our proven P2O5 trace moisture measurement method:

  • Integrated sampling system with rotameter (standard calibration for air)
  • No drifting of the measurement results
  • Robust, durable sensor element
  • No drifting of the measurement results
  • No follow-up costs as maintenance-friendly
  • Offset | Span | Bases possible via digital interface
  • Dry to wet: T90 < 5 sec; Wet to dry: T90 < 15 min
  • Significantly higher resolution with 0.01 ppmV in the range 0-500 ppmV and 1 ppbV in the range 0-20,000 ppbV
  • Analogue output and digital communication via RS232
  • Limit and status relays
  • 1.8" TFT display
  • Ex(i) version on request

 

Technical Specifications

Product: AQUATRACE® easyLINE - Trace moisture analyzer with integrated sampling system
 

Performance Specifications

Measuring range: 0-500 ppmV or alternatively 0-20,000ppbV, on request ranges deviating from the standard available
Accuracy: PPM: +/- 2 % of the display value, min. 0.4 ppmV | PPB: +/-2% of the display value, min. 10 ppbV
Response time: Dry to moist: T90 < 5 sec; Wet after dry: T90 < 15 min
Repeatability: PPM: +/- 0.01 ppm | PPB: +/- 5 ppb
Calibration: traceable calibration certificate
Display 1.8" TFT color display (output ppmV and °C dp)
 

Electrical Specifications

Output: 4-20 mA or 0-10 V, RS232 (9600, 8, n, 1)
Exit: ppm, ppb (°C Tdp via digital interface)
Scaled range of analog output: PPM: 0-500 Vppm | PPB: 0-20,000ppb
Power supply: 19-26 V DC
Power consumption: maximum 1.7 A
Conformance CE, EN 61000-4-2 EN 61000-4-3, EN 61000-4-4, EN 55011, EN ISO 12100
 

Service conditions

Operating temperature: -10 to 60°C
Storage: -10 to 60°C
Flow: Ensuring a constant flow rate!; Standard chamber: 20Nl/h
Working pressure: 0-10 bar absolutely | Special variant up to max. 30bar possible
 

Mechanical Specifications

Protection: IP65
Housing material: Electronics: Aluminium, anodized
Sensor: Pt-interdigital structure on ceramic carrier material (round)
Process connection: KF40, KF25 flange, G1/2 IG or measuring chamber with G1/8" female thread at the gas input and gas output
Electrical connections: M12 8-pin, 3m cable with open cable ends
Weight:

0.7 kg-1.3kg

 

Gas connections

Gas input compression fitting for 3mm stainless steel tube on the back
Gas output compression fitting for 3mm stainless steel tube on the back

 

 

 