The AQUATRACE® easyLINE is our new, particularly inexpensive, P2O5 trace moisture analyzer for ppmV | ppbV for inert and combustible gases with integrated sampling system. The system measures the volume-related moisture content in the gas and outputs it as ppmV and also approximately converts this value into the corresponding dew point temperature. Suitable for process and industrial moisture measurement, the measuring system is also equipped with limit alarm relays and status relays. The 1.8" TFT display visualizes the measured value. If the range is exceeded or not reached, the display color changes to flashing red. In addition to an analogue output 4-20mA, the analyzer also has an RS232 interface, which can be used to read out the measured value, but also for adjustment to local conditions.
Particularly suitable for the following sectors:
- Glove boxes | insulators
- Vacuum technology
- Plasma, sintering and soldering furnaces
- Technologies with protective gases (welding; soldering; thermal treatments, etc.)
- Coating plants
Further advantages of our proven P2O5 trace moisture measurement method:
- Integrated sampling system with rotameter (standard calibration for air)
- No drifting of the measurement results
- Robust, durable sensor element
- No drifting of the measurement results
- No follow-up costs as maintenance-friendly
- Offset | Span | Bases possible via digital interface
- Dry to wet: T90 < 5 sec; Wet to dry: T90 < 15 min
- Significantly higher resolution with 0.01 ppmV in the range 0-500 ppmV and 1 ppbV in the range 0-20,000 ppbV
- Analogue output and digital communication via RS232
- Limit and status relays
- 1.8" TFT display
- Ex(i) version on request
Technical Specifications |
|
| Product: | AQUATRACE® easyLINE - Trace moisture analyzer with integrated sampling system |
Performance Specifications |
|
| Measuring range: | 0-500 ppmV or alternatively 0-20,000ppbV, on request ranges deviating from the standard available |
| Accuracy: | PPM: +/- 2 % of the display value, min. 0.4 ppmV | PPB: +/-2% of the display value, min. 10 ppbV |
| Response time: | Dry to moist: T90 < 5 sec; Wet after dry: T90 < 15 min |
| Repeatability: | PPM: +/- 0.01 ppm | PPB: +/- 5 ppb |
| Calibration: | traceable calibration certificate |
| Display | 1.8" TFT color display (output ppmV and °C dp) |
Electrical Specifications |
|
| Output: | 4-20 mA or 0-10 V, RS232 (9600, 8, n, 1) |
| Exit: | ppm, ppb (°C Tdp via digital interface) |
| Scaled range of analog output: | PPM: 0-500 Vppm | PPB: 0-20,000ppb |
| Power supply: | 19-26 V DC |
| Power consumption: | maximum 1.7 A |
| Conformance | CE, EN 61000-4-2 EN 61000-4-3, EN 61000-4-4, EN 55011, EN ISO 12100 |
Service conditions |
|
| Operating temperature: | -10 to 60°C |
| Storage: | -10 to 60°C |
| Flow: | Ensuring a constant flow rate!; Standard chamber: 20Nl/h |
| Working pressure: | 0-10 bar absolutely | Special variant up to max. 30bar possible |
Mechanical Specifications |
|
| Protection: | IP65 |
| Housing material: | Electronics: Aluminium, anodized |
| Sensor: | Pt-interdigital structure on ceramic carrier material (round) |
| Process connection: | KF40, KF25 flange, G1/2 IG or measuring chamber with G1/8" female thread at the gas input and gas output |
| Electrical connections: | M12 8-pin, 3m cable with open cable ends |
| Weight: |
0.7 kg-1.3kg |
Gas connections |
|
| Gas input | compression fitting for 3mm stainless steel tube on the back |
| Gas output | compression fitting for 3mm stainless steel tube on the back |