The AQUATRACE® ATT520V is our new P2O5 trace moisture analyzer for ppmV | ppbV for inert and combustible gases. Our ATT520 measures the volume-related moisture content and approximately converts this value into the corresponding dew point temperature. Suitable for process and industrial moisture measurement, the measuring system is also available with a limit alarm relay and status relay. The large 1.8" TFT display (optionally available) visualizes the measured value in an easy-to-read manner. If the range is exceeded, the display color changes to flashing red. In addition to a 4-20mA analog output, the analyzer also has an RS232 interface, which is used to read out the measured value but also for adjustment to local conditions. A watchdog (optional) monitors the electronics and a status relay (optional) sends the functional status to the control room.
5 practical standard variants available:
• By-Pass-Solution with 20Nl/h standard measuring chamber
• DN25KF/DN40KF flange adapter for inline-measuring
• G1/2-Thread for inline-measuring
• electronics with separate sensor (in the above variants: standard chamber | DN25KF | DN40KF | G1/2")
• Intrinsically safe variant: Ex II 2G Ex Ia IIC T4 Gb (using a ZENER-barrier)
Particularly suitable for the following sectors:
• glove-boxes | isolators
• vacuum technology
• Plasma, sintering and soldering furnaces
• Technologies with protective gases (welding; soldering; thermal treatments, etc.)
• coating plants
Further advantages of our proven P2O5 residual moisture measurement method:
• No drifting
• No follow-up costs as maintenance-friendly
• Offset | Span | Bases possible via digital interface
• Dry to wet: T90 < 5 sec; Wet to dry: T90 < 15 min
• Significantly higher resolution of 0.01ppm in the 0-500ppm range and 1ppb in the 0-20,000ppb range
• Further improved response
• Analogue output and digital communication via RS232
• Optional: limit value relay, watchdog, large 1.8" TFT display, Ex(i) version
Technical Specifications |
|
| Product: | AQUATRACE® ATT520V - Trace Moisture Analyzer |
Performance specification
|
|
| Measuring range: | 0-500 ppmV or 0-20,000ppbV, other ranges on request |
| Accuracy: | PPM: +/- 2 % of the display value, min. 0.4 ppmV | PPB: +/-2% of the display value, min. 10 ppbV |
| Ansprechzeit: | Dry to wet: T90 < 5 sec; Wet to dry: T90 < 15 min |
| Repeatability: | PPM: +/- 0,01 ppm | PPB: +/- 5 ppb |
| Calibration: | traceable calibration certificate |
| Display | 1.8" TFT color display |
Electrical Specifications |
|
| Output: | 4-20 mA or 0-10 V, RS232 (9600, 8, n, 1) |
| Exit: | ppm, ppb (°C Tdp via digital interface) |
| Scaled range of analog output: | PPM: 0-500 Vppm | PPB: 0-20.000ppb |
| Power supply: | 19-26 V DC |
| Power consumption: | maximal 1,7 A |
| Conformance | CE, EN 61000-4-2 EN 61000-4-3, EN 61000-4-4, EN 55011, EN ISO 12100 |
Service conditions |
|
| Operating temperature: | -10 to 60°C |
| Storage: | -10 to 60°C |
| Flow: | Ensuring a constant flow rate!; Standard chamber: 20Nl/h |
| Working pressure: | 0-10 bar absolutely | Special variant up to max. 30bar possible |
Mechanical Specifications |
|
| Protection: | IP65 |
| Housing material: | Electronics: Aluminium, anodized |
| Sensor: | Pt-interdigital structure on ceramic carrier material (round) |
| Prozessanschluss: | KF40, KF25 flange, G1/2 IG or measuring chamber with G1/8" female thread at the gas input and gas output |
| Electrical connections: | M12 8-pin, 3m cable with open cable ends |
| Weight: |
0.7 kg-1.3kg |