Smart Moisture Control 0-2,000 ppmV | 0-20,000 ppbV | optionally with FlowControl
Continuous measurement of trace moisture for corrosive and neutral gases for integration into your processes
|
Technical specifications
Display |
0-6,000ppmv |
Accuracy |
0,4ppmv or 5% of measuring value(0-2,000ppm) 0,4ppm or 2% of measuring value(0-500ppm) 10% of measuring value(0-20,000ppb) |
Resolution |
0,01ppm(ppm Measuring range) or 1ppb(ppb Measuring range) |
Reaction time |
< 1 s |
Action time T90 (high) |
< 5 s |
Action time T90 (down) |
< 15 min |
Diagnostic function |
Flow monitoring, self-diagnosis of sensor and controller, heartbeat monitoring |
Alarm logger |
Internal flash memory, up to 6000 alarms |
Analog output |
4 - 20mA, maximum load 500Ω (galvanically separated) |
Relay output |
Relay (2A, 230V AC), system alarm, galvanically separated |
Control function |
Optional timer controller, PID analog controller, PWM controller |
Calibration / adjustment |
Expert calibration function, multi-point calibration function up to 9 points |
Communication |
RS485 MODBUS RTU, HART, Foundation Fieldbus FF, PROFIBUS PA, PROFIBUS DP, MODBUS TCP/IP, etc |
Power |
80~264V AC,1A or 19~28V DC,3A |
Electrical protection |
EMI / RFI CEI-EN55011 – 05/99 |
Ambient temperature |
-15 ~ 60°C |
Storage and transport temperature |
-25 ~ 70°C |
Gas flow |
20Nl/h or 100Nl/h |
Process pressure (max.) |
3Bar(PVDF) or 10Bar(Stainless steel) |
Sample gas temperature |
5~65°C |
Process connection |
1/4"NPT thread or KF40 Flansch |
Pipe connection |
6mm |
Leak rate |
< 5x10-8 mbar x l / s-1 |
Order key
SMC- |
MT |
X |
X |
X |
X |
X |
||
|
MT — Trace moisture analyzer for gases |
|||||||
|
|
50-4.3 "industrial color-fast touch screen, 480 x 272, 4-20mA output, relay output |
||||||
|
|
70—7 "capacitive industrial color touchscreen, 1060 x 600.4 20 mA output, relay output |
||||||
|
|
|
W-Wall mounting |
|||||
|
|
|
W (Ex) - Wall mounting with Ex d II CT4 housing |
|||||
|
|
|
R—Rack(19") |
|||||
|
|
|
T — Portable |
|||||
|
|
|
A—Portable (including battery) |
|||||
|
|
|
|
1—Singel channel (standard) |
||||
|
|
|
|
2—Two channel |
||||
|
|
|
|
2O—Two-channel, residual moisture and O2 measurement (0 ~ 1000ppm) |
||||
|
|
|
|
|
D—19 ~ 28V DC |
|||
|
|
|
|
|
A—80 ~ 264V AC |
|||
|
|
|
|
|
|
0—RS485, MODBUS RTU |
||
|
|
|
|
|
|
1—HART |
||
|
|
|
|
|
|
2—Profibus-PA |
||
|
|
|
|
|
|
3—Profibus-DP |
||
|
|
|
|
|
|
4—Foundation FieldBus |
||
|
|
|
|
|
|
X—customer special (on request) |
||
|
|
|
|
|
|
|
FI — flow meter Namur input |
|
|
|
|
|
|
|
|
C1 — PID control (AO3) |
|
|
|
|
|
|
|
|
C2 — timer control (DO2) |
|
|
|
|
|
|
|
|
C3 — PWM control (DO2) |
Measuring cell selection
ATM |
X |
X |
X |
M8 |
X |
|
|
BP—PVDF |
|||||
|
BS—SS304 stainless steel |
|||||
|
|
G — glass sensor (bifilar platinum wire on glass rod) |
||||
|
|
K—Pt thick film on a ceramic carrier |
||||
|
|
CerMet — Pt thick film on a ceramic carrier for corrosive gases |
||||
|
|
|
20 — Flow measuring cell 20Nl / h |
|||
|
|
|
100 — flow measuring cell 100 Nl / h (a little faster with the reaction) |
|||
|
|
|
|
M8 — connection |
||
|
|
|
|
|
Ex — intrinsically safe Ex ia |