Trust is good, but measuring is even better!
Designation: precise and maintenance-free

PM 2000+ The paramagnetic oxygen analyzer in range from 0,1% up to 100% - precise and maintenance-free

Brief description

The PM 2000+ is a precise oxygen analyser for continuous monitoring purposes at an amazing price. Build in a modular housing system, with a modern micro-controller technology, it is specially designed for process and ambient air measurements. The analyser is equipped with an auto-calibration functionality and a self-diagnosis capability and has an RS232 interface. With the programmable auto calibration function it is capable to fulfil a fully automatic calibration by means of the integrated relays. The measuring unit is thermostat temperature controlled to 55°C. The operation and parameterisation is carried out by means of a user friendly 4 keys and a 16-digit LCD display and also through RS232 interface

Measuring principle

The measurement is based on the paramagnetic characteristic of oxygen. It generates a partial pressure within a strong and non-homogeneous magnetic field which moves a rotatable glass dumbbell within the measuring cell. This small rotation is measured by the projection of a light source on a photodiode via a small mirror on the dumbbell. A small current through a wire around the dumbbell forces the dumbbell to its initial position. This current is amplified and is directly proportional the oxygen concentration.

PM 2000+
19” 3 HU rack or table model with or without sample test gas
filter and flow display on the front panel (option). Single or multiple channel versions.
WxHxD: 482 x 133 x 245 mm


IP protection class: 20

PM 2100+

1/2 19” portable model optional with filter on the rear. WxHxD: 235 x 155 x 280 mm

IP protection class: 52

PM 2200+

wall-mounted housing Rittal, Single or multiple channel versions.

WxHxD: 380 x 410 x 210 mm

IP protection class: 52

 

Specifications

Measurement range output

free settable by input of parameters between 0- 100% O2

Measurement range

Standard range 0-25%, 0-100%, others on request

Measurement signal

4-20mA or 0-20mA (max. apparent ohmic resistance 500 Ohm)

Status output

2 alarm relays, 1 malfunction relay

Output connection

pump relay, maintenance

sample gas relay, zero gas relay, test gas relay for the autocalibration

Display

LCD digital multifunction display, indication of measured value:

100.0 %O2; Flow 99l/h, alarms, malfunction, parameters, total 16 digits

Options

 

 

·        pressure compensation (electronic or backpressure regulator)

·        test gas pump

·        sample gas filter

·        external flow display (rotameter)

·        flow sensor with alarm setting

·        NDIR sensors for CO2, CH4 and CO

 

Design

 

Materials of gas conducting parts

PVDF, glass, steel 1.4571, gold, viton, platinum-iridium, epoxy resin, nickel

Gas connections

PVDF bulkhead connection, for hose with inside ø 4mm

Power supply

 

Voltage

100 - 240VAC 50/60Hz

Operating conditions

 

Flow

10-90 l/h

Gas conditioning

necessary for humid and/or corrosive gases, pre-filter required

Operating gas pressure

10-1000 hPa (0.01-1bar)

Operating temperature

5-45°C

Calibration

2-point calibration: with gases as desired, menu-controlled, time controlled.

fully automatic AUTOCAL or semiautomatic calibration

Storage and transport temperature

-25°C to +65°C

Relative humidity

0-75% RH

Background gas influence

slight (for guideline data see operating instructions)

Measuring details

PM 2000+

Detection limit

0,01 % O2

Repeatability

<= ± 0,03 % O2 (time base for gas switch >= 5 min)

Zero point drift

<= ± 0,1 % O2 / week (offset)

may be higher during the first days after putting into operation or after longer period of storage or transport

Temperature influence at zero

< ± 0,05 % O2 /°C

Temperature influence span

< ± 0,20 % of measured value / °C

Pressure influence on zero

no influence

Pressure influence span

1% air pressure change causes 1% change in reading without backpressure regulator (option) or pressure compensation (option)

Flow error

<= 0,1 Vol.-% O2 within 10…90 l/h

with the in-build flow regulator (option)

T90-time

<= 6s at 90 l/h and gas change from nitrogen to air

 

« Back to overview