DKS GmbH - Produkte
Trust is good. Measuring is better!
Designation: Fast Thermal Conductivity Analyzer

The FTC300 is a thermal conductivity analyzer with outstanding properties for measuring tasks in process analytics. The device is characterized by high measurement accuracy and stability. It has a comfortable operating menu that allows you to adjust, parameterize and configure without much effort. In addition, operator and data access via an RS232 interface are possible, as well as various analog signal inputs and outputs. The robust design of the FTC300 is designed in protection class IP65 and enables operation at ambient temperatures of -20 ° C to + 50 ° C and is pressure-resistant up to 20 bar. For the measurement of hydrogen in nitrogen, the smallest measuring range is 0 to 0.5 vol.%.

The thermal conductivity analyzer FTC300 can be set up in such a way that switching between different measuring tasks is possible. Concentration determination in up to 16 different binary gas mixtures is possible. This option can be ordered under the keyword “Multi Gas Mode”.

The FTC300 provides a highly effective routine for cross-sensitivity compensation, which can be flexibly adapted to the respective measurement task. It requires as an input external voltage signal, which represents the concentration of the cross-sensitivity causing gas components.

Specification

General:
Measurement process Thermal conductivity
Sampling Analyzer extractive
Connection 6mm pipe
Dimensions (WxHxD in mm) 145x80x85
Protection class IP 65
Pressure-proof to 20 bar absolute
Power supply 18V to 36V DC / 700mA
Ambient temperature range -20°C to 50°C
Weight to 1800g
T90 time at 60l/h < 1sec
Communication:
RS232 Yes
Current output 1x, 0/4-20 mA
Voltage output 2x, 0-10V
Display Yes
Relays 3x
Service program SetApp 2.0 Yes
Calibration on site Yes
Options
Flow measurement Yes
Flow monitor Yes
Infrared measurement
Multi Gas Mode Yes
Protection against corrosion Yes
Protection against condensate and dust Yes
Suitable for inflammable gases Yes
Moisture measurement Yes
Cross-sensitivity compensation Yes, external signal required
Specifications gas analysis:
Noise < 1% of smallest range
Drift at zero point per week < 2% of smallest range
Repeatability < 1% of smallest range
Non-linearity < 1% of range
Measuring error with ambient temperature change per 10°K < 1% of smallest range
Flow influence between 60l/h and 90l/h per 10l/h < 1% of smallest range
Fault with measurement gas change (Pabs > 800 hPa) per 10 hPa < 1% of smallest range